HTF-1100

HTF-1100

HTF-1100 consists of a temperature control unit, an alumina heater and a slide on the front panel to insert and extract easily the sample holder from the furnace.

HTF-1100 consists of a temperature control unit, an alumina heater and a slide on the front panel to insert and extract easily the sample holder from the furnace.

ITS

ITS

Intermediate Temperature System (ITS) is a compact temperature control unit based on the Peltier effect.

It enables accurate temperature control from –35 °C to +150 °C. The ITS is designed to be used with a CESH sample holder provided with the unit.

Temperature range: -35 °C to +150 °C
Temperature accuracy: ±0.3 °C
Computer Interface: USB 2.0
Power consumption: 250 W
Dimension: 400 x 313 x 385 mm (L x W x H)
Weight: 8 kg

MTZ-35

MTZ-35

Exploring new frontiers of impedance testing

The Bio-Logic MTZ-35 impedance analyzer has the specifications and features required to address the broad scope of applications in the materials research field.

Frequency Range: 10 μHz to 35 MHz
Voltage range DC: ±100 μV to ±10 V
Voltage range AC: 100 μV to 10 V

Measurement range
Inductance: 10 nH to 10 kH
Capacitance: 1 pF to 1000 μF
Resistance: 1 mΩ to 500 MΩ
Basic accuracy: 0.1%

MCS 10

MCS 10

µS to S from -40 °C to +150 °C

Based on our expertise in impedance spectroscopy, Bio-Logic has developed the MCS 10, a unique integration of a frequency response analyzer and a temperature control unit, enabling mutli-sampling at various temperatures.

Conductivity ranges: 6 ranges from 2 μS/cm to 0.2 S/cm
Temperature range: -40 °C to +150 °C
Temperature Accuracy: +/-0.2% (typ.) +/-0.1 °C
Number of samples:10 slots for 12 mm vials
Computer connection: USB 2.0
Cell type: Parallel plates platinized platinum on glass holder (x10)